ppb level trace NOx gas analyzer
Product Features
1.touch color screen, display rich content, convenient shortcut menu query;
2.using a single reflex chamber and a single photomultiplier tube, NO and NOx are measured by automatic cycle conversion;
3.stable operation, high precision, good repeatability, simple operation;
4.automatically store calibration data, alarm information, easy to check the operation status;
5.support one-click query of historical data, run data curve personalized display;
6.support remote software system upgrade, safe and convenient;
7.with startup self-test and self-diagnosis function, convenient for users to quickly locate and judge faults;
Product Parameters
Model No. | TY-AQMS -42 |
Analytical method | chemiluminescence method |
Measuring range | 0-500ppb |
Zero noise | ≤1ppb |
range noise | ≤5ppb |
Minimum detection limit | ≤2ppb |
indicates error | ±2%F.S. |
Response time (rise/fall) | ≤300s |
Voltage stability | ≤±1%F.S. |
Flow stability | ≤±10% |
Influence of ambient temperature change | ≤3ppb/℃ |
Conversion efficiency | > 96% |
Influence of interfering components | 2.5%H2O: ≤±4%F.S. 1μmol/molNH3: ≤±4%F.S. 0.2μmol/molO3: ≤±4%F.S. 0.5μmol/molSO2: ≤±4%F.S. |
Concentration deviation between sampling port and calibration port | ≤1% |
Unattended hours | Long-term zero drift: ≤10ppb/7d Long-term range drift: ≤±20ppb/7d |
Average days between failures | ≥7 days |
Application Scenario

Environmental air monitoring

industrial park leakage monitoring

Regional atmospheric background monitoring

air and pollution source emission impact monitoring

Determination of gas content in various combustion tests in the laboratory

Pollution control assessment project
INTRODUCTION
Tianjin ShareShine Technology Development Co., Ltd. is a high-tech enterprise with independent innovation as the driving force and scientific and technological development as the core, which closely integrates "production, learning, research and application". The company has international first-class level in the field of spectrum detection technology and environmental monitoring technology. The company's main business covers environmental online monitoring instruments, environmental Internet of Things system solutions and intelligent operation and maintenance services.
The company has always adhered to independent innovation, focusing on the research and development and manufacturing of high-end environmental monitoring equipment. Widely used and service in power plants, sewage treatment plants, pharmaceutical, printing, textile, petrochemical, gas safety, underground pipe network, safety protection, precision manufacturing, mining and metallurgy, university research, ambient air and water treatment, light industry and electronics industry, water supply and drinking water distribution network, food and beverage, hospitals, hotels, aquaculture, new agricultural cultivation and biological fermentation Crafts and other industries.

Head Office

Optical Lab

R&D

Production Workshop

Chemistry Lab